[email protected]

Home   |   Imaging & Digital Dentistry

Imaging & Digital Dentistry

Digital Sensor Positioner with Aiming Ring

  • Type: Paralleling technique
  • Fits: Multiple sensor sizes
  • Sterilisation: 121 C autoclavable
  • Colours: Green, blue, yellow
  • Product description: Paralleling-technique positioner for wired digital sensors. Adjustable to fit multiple sensor sizes, colour coded, autoclavable at 121 C.
  • INQUIRY

A sensor positioner exists to make one thing repeatable: the angle between the sensor, the tooth and the cone. When that angle is right, the image is a parallel projection of the tooth and the buccal and lingual cusps sit on top of each other. When it is off, the crown looks longer or shorter than it is, and the same tooth measured twice gives two different numbers.

This positioner is built for wired digital sensors. It is adjustable so one kit covers the sensor sizes a practice actually owns, and it is autoclavable so it can go back on the tray after every patient rather than being wiped down and reused.

How the assembly works

  • The sensor cradle clamps the sensor and the cable exit, so the cable does not pull the sensor out of position mid-exposure.
  • The arm is adjusted in one direction only. Single-axis adjustment holds its setting; a ball joint looks more flexible on paper but drifts under cable tension.
  • The aiming ring sits at a fixed distance from the sensor. Align the cone to the ring and the beam is perpendicular to the sensor by geometry, not by eye.
  • Colour coding (green, blue, yellow) lets one kit serve anterior, posterior and bitewing positions without hunting for the right block.

Sizes and compatibility

The cradle is adjustable rather than fixed, which is the point: a practice running two sensor generations, or a dealer supporting several customer brands, does not need a separate holder for every size.

ItemDetail
ApplicationWired digital sensors (RVG)
AdjustmentSingle-direction, tool-free
Sensor fitAdjustable cradle, multiple sizes
SterilisationAutoclave up to 121 C
PositionsAnterior, posterior, bitewing
Supplied asComplete kit with rings and bite blocks

Setting it up correctly

  • Fit the sensor into the cradle before the patient bites. Pushing a sensor into a bitten block moves the tooth.
  • Seat the bite block on the incisal or occlusal surface the patient actually closes on, then ask them to close gently and hold.
  • Check the ring is square to the cone head before exposing. A ring that is visibly oval means the cone is off-axis.
  • Autoclave the metal and high-temperature parts; wipe the sensor and cable with a compatible disinfectant, never immerse them.

What to check before ordering

Two things decide whether a positioner works in a given practice. The first is sensor thickness: a cradle sized for a thin sensor will not close on a thicker one, and forcing it cracks the housing. The second is the cable exit direction, because a cradle that routes the cable across the bite block will not seat.

Send us the sensor model and the sizes you use and we will confirm the fit before shipping. Kits are supplied boxed with the bite blocks and rings included; spare rings and blocks are available separately for practices that lose them.